Application
ABX Pentra XL 80
Overview
• CBC with 5-part Diff including flagging for immature cells
• Advanced technologies for mixing and sample analysis ensuring accurate results
• Processes up to 80 samples/hr and provides results in approximately 60 seconds
• Autoloader holds up to 100 samples with random continuous access
• STAT mode for critical samples
Features
• 20 parameters
• Whole blood micro-sampling (CBC: 35 µL, Plus Diff 53 µL)
• Cytochemistry, Impedance (real cell volume measurement) & Optical (analysis of the internal cellular structure by measuring light absorbance) - DHSS* Technology
• Perfect homogenization of blood samples with reagents - MDSS** Technology
• Automatic Sample Re-run
• Reagents: Only 4 reagents plus cleaner
• Perfect differentiation of the 5 WBC sub-populations with DHSS* Technology
• 3 histograms for RBC, BAS/WBC and PLT together with the 5 DIFF Matrix
• Basophils counted through specific channel
• Customized Dilution Ratio (CDR)
• Integrated Workstation and Validation Station
* DHSS : Double Hydrodynamic Sequential System
** MDSS : Multi-Distribution Sampling System
Manufactured by HORIBA ABX SAS
Specifications
Technical
Weight | 122 lb / 55 kg | ||
Power supply | 100 V to 240 V (± 10%) - 50 Hz to 60 Hz | ||
Dimensions | 32.3 x 22.4 x 21.5 in / 82 cm x 57 cm x 54 cm. | ||
Printer | Digital Laser | ||
Throughput | Up to 80 samples/hour | ||
Data Processing | Color LCD touch screen: 12 in PC onboard: Celeron 566 MHz Windows XP Prof.TM RAM (256 MB), Hard disk (10 GB mini) Floppy disk & CD ROM reader RS 232C, 2 X USB1 User defined flagging limits Transmit patient files & QC to LIS Uni-directional & bi-directional connections ASTM protocol inside |
| |
Memory capacity | 10 000 results + Graphics | ||
Quality Control Management | 24 selectable QC files XB: 60 operator selectable files with statistics (20 results per file) Within run Levey-Jennings graphs |
| |
Logs | Reagents, Quality Control, Calibration, Maintenance, Blank Cycle, Patients Data, Settings, Information, Technology Communication, Errors, by date. |
|
|
Patient reports Management | Delta check Archived results (Matrix, curves, data) Manual input |
|
|
Samples
80 samples/hour (open or closed tube) | |
|---|---|
Parameters (26) | WBC, NEU# & NEU%, LYM# & LYM%, MON# & MON%, EOS# & EOS%, BAS# & BAS%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV |
Graphs | High-definition LMNE matrix |
Blood Volume | CBC mode: 35 µL |
Stability | 48 hours post-draw stability |
Operating sample mode | Autoloader 10 racks (100 tubes) capacity |
Identification | Alphanumeric. Bar-code reader (optional) and internal bar-code scanner |
Results
Parameters | Precision (%CV) | Linearity Limits | Linearity Limits with CDR* |
|---|---|---|---|
WBC | < 2.0 | 0-120 x 10³/µL | 120-360 x 10³/µL |
RBC | < 2.0 | 0-8 106/µL | 0-8 X 106/µL |
HGB | < 1.0 | 0-24 g/dl | 0-24 g/dL |
HCT | < 2.0 | 0-67% | 0-67% |
PLT (whole blood) | < 5.0 | 0-1900 x 10³/µL | 1900-3800 x 10³/µL |
PLT (PLT concentrate) | < 5.0 | 0-2800 x 10³/µL | 2800-5600 x 10³/µL |
*CDR: Customized Dilution Ratio