On top of using a double moulding injection process integrating polycarbonate resin and TPE to create an eco-friendly, shock & scratch resistant, non-slip body, several improvements and enhancements have been made to the latest HORIBA LAQUA handheld meters allowing sustained periods of operations in challenging outdoor conditions and making the 200 series a compact yet powerful versatile meter.
LAQUAact PC110
Overview
Features
- Dual channel, switchable display
- Automatic pH calibration up to 5 points using NIST, USA or Custom pH buffers
- 0.01 pH resolution
- Offset and slope display
- Automatic / Manual temperature compensation and temperature calibration function
- Automatic and manual conductivity calibration up to 4 points
- Adjustable reference temperature and temperature coefficient
- Selectable cell constant and measurement unit
- Conductivity and total dissolvd solids (TDS) conversion
- Four (4) TDS calibration curves for wide range of applications: EN27888, Linear, 442, NaCl
- Two (2) salinity calibration curves: NaCl, Seawater
- Auto log data; Built-in memory can store up to 1000 data
- Auto shut-off (programmable: 1 to 30 minutes)
- Auto-Hold / Auto-Stable measurement
- Calibration alarm (programmable: 1 to 400 days)
- Built-in electrode holder allows 2 electrodes to be attached on both sides of the meter
- Large backlit LCD: 52mm x 64mm
- RS232 output for connecting to computer or printer
- Casing is IP67 waterproof (for 30mins immersion in 1m depth of water) and dustproof
- Powered by 2 x AAA (LR03) or Ni-MH rechargeable batteries or power adaptor
- Packaged in new molded carry case
- 2 years meter warranty
Meter Kit Available
PC110-K
The appearance of items in the picture are subject to change in each region.
Package Details:
- LAQUAact PC110 Dual channel pH/ORP/Cond/TDS/Res/Sal/Temp meter, built-in electrode holder, 2 x AAA batteries
- Instruction manual
- 9651-10D plastic-body, gel-filled pH electrode with integrated temperature sensor
- 9382-10D plastic-body, Titanium/Platinum black, conductivity probe (k=1.0) with integrated temperature sensor
- pH 4.01 and 7.00 buffers, 1413µS/cm and 12.88 mS/cm conductivity solutions (60ml each)
- Carry case
PC110-KN: Kit with pH 4.01 and 6.86 NIST buffers (60ml each) is available upon request.
Manufactured by HORIBA Advanced Techno
Specifications
Model | LAQUAact PC110 Dual Channel pH/ORP/Cond/TDS/Res/Sal/Temp |
---|---|
pH Range | -2.00 to 16.00 pH |
Resolution | 0.01 pH |
Accuracy | ± 0.01 pH |
Calibration Points | Up to 5 |
Buffer Options | USA, NIST, Custom |
ORP Range | ± 2000 mV |
Resolution | 1 mV |
Accuracy | ± 1mV or 1% of reading (whichever is greater) |
Conductivity Range | ...µS/cm to 2000 mS/cm |
Resolution | 0.05% F.S. |
Accuracy | ±0.6% F.S.; ±1.5% F.S. > 18.0 mS/cm |
Reference Temperature | 15 to 30°C (adjustable) |
Temperature Coefficient | 0.0 to 10.0% (adjustable) |
Cell Constant | 0.1, 1.0, 10.0 (selectable) |
Calibration Points | Up to 4 (Auto / Manual) |
Unit Setting | µS/cm, mS/cm, S/m (Auto-ranging) |
TDS Range | ...ppm to 1000 ppt (TDS factor=0.5) |
Resolution | 0.01 ppm / 0.1 ppt |
Accuracy | ±0.1% F.S. |
Calibration Curves | EN27888, 442, Linear, NaCl |
Resistivity Range | 0.000 Ω•cm to 200.0 MΩ•cm |
Resolution | 0.05% F.S. |
Accuracy | ±0.6% F.S.; ±1.5% F.S. > 18.0 mΩ•cm |
Salinity Range | 0.0 to 100.0 ppt / 0.00 to 10.00% |
Resolution | 0.1 ppt / 0.1 % |
Accuracy | 0.2% F.S. |
Calibration Curves | NaCl, Seawater |
Temperature Range | -30.0 to 130.0°C |
Resolution | 0.1°C |
Accuracy | ± 0.4°C |
Calibration Option | Yes |
Memory Data | 1000 |
Auto Log Data | Yes |
Real-time Clock | Yes |
Date / Time Stamp | Yes |
Auto-Shut Off | Yes (Programmable: 1 to 30 minutes) |
Auto-Hold / Auto-Stable | Yes |
Offset / Slope Display | Yes (Independent acid and alkaline slopes depending on calibration) |
Calibration Alarm | Yes (Programmable: 1 to 400 days) |
Electrode Status | On screen display |
Diagnostic Messages | Yes |
Display | Custom Backlit LCD / Dual Channel / Switchable Display |
Inputs | 2 x BNC, 2 x phono, DC sockets |
Output | RS232 |
IP Rating | IP67, shock-resistant |
Power Requirements | 2 x AAA (LR03) or Ni-MH rechargeable batteries, or power adaptor |
Battery Life | >200 hours |
Downloads
HORIBA Offices
Asia & Oceania
Europe, Middle East & Africa
Northampton
HORIBA UK Limited
Northampton Office
work
Kyoto Close,Summerhouse Road,
Moulton Park
Northampton NN3 6FL UK
Moulton Park
work
+(44) 1604 542 600
work
fax
+(44) 1604 542 699
waterquality@horiba.com
Americas
Irvine, California
HORIBA Instruments Incorporated
Head Office
work
9755 Research Drive
Irvine California 92618 USA
work
+1 800 446 7422
work
fax
+1 949 468 1790
labinfo@horiba.com